2001B / 2001C

 

Características Técnicas

(English) 

 

 Inicio
 

Input Circuit

Protect the instrument from tilting and possible vibration or shock

Patient leads

Standard 12 leads manual / auto

Patient current leakage

< 10 uA

Input impedance

>= 50 M ohms

Calibrating voltage

1mV +/- 3 %

A/D Exchange

12 bit

Frequency response

0.05 Hz – 150 Hz

Time constant

> 3.2 seconds

Common mode

> 80 dB

> 100 dB (rejection controller)

Rejection rate Electromyoplasmy interference

35 Hz (-3dB)/ 25 Hz (-3dB) 001

Sensitiviti setting

1/2 ,1, 2, (cm/mV), exchange tolerance <= 5%

Recording mode

Thermal array printing system

Printing speed

25, 50 mm/ second  +/- 3%

Width of ECG paper

50mm, 20m high-speeding thermal printing paper

Input circuit

<= 50nA

Safety standard tape

IEC 1/ CF with protecting circuit for vibration or shock

Power supply

AC: 220 /110V, 50/60 Hz

DC: 12 V rechargeable battery

Skin voltage

+/- 300 mV

Baseline drift

< 0.3 mm

Noise level

< 15 u Vp-p

Dimensions

(LxWxH) 31 x 23 x 7 Cm / 12,2 x 9 x 2,7 inches

Dimension of package

(LxWxH) 40 x 28 x 16 Cm / 15,7 x 11 x 6,3 inches

Net weight

2,7 Kg / 6 Lbs

Gross weight

4,5 Kg / 10 Lbs